MiniCAT™ User Manual v5.2
© 2014 Xoran Technologies LLC
10.3 X-Ray Specifications
X-Ray Specifications
Source Block Manufacturers
Source-Ray, Inc., Imaging Sciences International, Inc.
(ISI)
Source Block Model
SB-125-7K
Tube Manufacturer
Superior X-Ray Tube Co.
Tube Model Number
SXR-130-15-0.5
Tube Type
Stationary anode
Nominal Tube Potential Range
120 - 125 kVp ±5% tolerance
Nominal Tube Current Range
r.m.s.
5 - 7 mA ±5% tolerance
Max. Tube Voltage
130 kVp (full wave rectified)
Max. Tube Current r.m.s.
8 mA
Target Material
Tungsten
Target Angle
15 degrees
Focal Spot Size
~0.5 mm (NEMA)
Minimum Inherent Filtration
2.0 mm Al @ 125 kVp
Rated Line Voltage
~120V AC, 50/60 Hz, 10A (r.m.s max line current)
(integrated into diagnostic system)
Range of Line Voltage
Regulation
2%
Maximum line current r.m.s
10A
Technique Factors that
Constitute the Maximum Line
Current Condition
125 kVp, 58.8 mAs
X-ray Generator Duty Cycle
Pulsed: 50% duty cycle, 11-30 ms pulse on time
Pulse Train: 30 pulses/sec (Hz)
Pulse Train Time: 40 seconds on, 10 minutes off
Rise Time (X-Ray Output)
Less than 5 ms @ 125 kVp 7 mA
measured @ 50% - 90 % of the voltage waveform.