128
MiniCAT™ User Manual v5.2
© 2014 Xoran Technologies LLC
Sinus 10s (150): 4.49 to 9.69
T Bone 20s (600): 3.54 to 7.91
11.9 Nominal Tomographic Section Thickness and Slice
Sensitivity Profile
Equipment and Scan:
Quality Control Phantom, Sinus 20s (600), Sinus 10s (300), Sinus 10s
(150), T Bone 20s (600)
Measuring Nominal Tomographic Section Thickness is obtained by finding the full-width half-
maximum (FWHM) resolution of the Slice Sensitivity Profile (SSP). The SSP is obtained by
scanning a small, sub-voxel bead in the QC module. Specifically, the custom QC module contains a
0.280 mm tungsten bead in the upper half of the phantom.
Note: See below picture as reference.