112
MiniCAT™ User Manual v5.2
© 2014 Xoran Technologies LLC
Manufacturer Recommended Frequency of QC Tests
Section
Test
Frequency
11.1
Laser Accuracy
Monthly
11.2
CT Number of Reference Materials
Monthly
11.3
Spatial Resolution
Monthly
11.4
CT Number and Standard Deviation of Water
Daily
11.5
Artifact Test
Daily
11.6
Uniformity Test
Monthly
11.7
Contrast Scale
Annually
11.8
Image Noise
Annually
11.9
Nominal Tomographic Section Thickness and
Slice Sensitivity Profile
Annually
Note:
The required frequency of QC tests may vary according to state regulations, accrediting
organizations, or your medical physicist.