8
MiniCAT™ User Manual v5.2
© 2014 Xoran Technologies LLC
Symbol
Description
Medical Grade Power Cord Symbol:
Non-Ionizing Electromagnetic Radiation
Symbol:
Serial Connection Symbol:
EC REP
RMS UK, Ltd., Somerset, BS48 4NU
United Kingdom
1.3 Electromagnetic Interference
The MiniCAT™ has been tested and found to comply with the limits for Class A equipment
pursuant to IEC 60601-1-2. These limits are designed to provide reasonable protection against
harmful interference in a commercial environment. The MiniCAT™ generates, uses and can radiate
electromagnetic energy. If the MiniCAT™ is not installed and used in accordance with these
instructions, it may cause harmful interference to surrounding equipment.
MiniCAT™ is intended for use in an electromagnetic environment specified below. The user of the
MiniCAT™ should ensure that it is used in such an environment.
Immunity / Test
Electrostatic Environment Guidance
Electrostatic Discharge
Compliant to +4 kV. Floors should be wood, concrete or ceramic tile.
If floors are synthetic, the relative humidity should be at least 30%.