106
MiniCAT™ User Manual v5.2
© 2014 Xoran Technologies LLC
Environmental Conditions
It is recommended that the MiniCAT™ is operated in a climate-controlled environment where the
ambient temperature does not exceed a range of 60 to 75 degrees Fahrenheit. An operating
environment where this suggested range cannot be regulated may reduce the expected life cycle of
the MiniCAT™ system. Changes in temperature may cause the system to require additional
calibration. Equipment not suitable for use in the presence of a flammable anaesthetic mixture with
air, oxygen or nitrous oxide.
Operating Requirements
60 to 80 degrees Fahrenheit
(16 to 26 degrees Celsius)
10% to 90% humidity
Transportation and Storage Requirements
14 to 104 degrees Fahrenheit
(-10 to 40 degrees Celsius)
10% to 90% humidity
Heat Dissipation
Approx. 500W / 1700 BTU/hr
Caution:
Allow system temperature to adapt to room temperature before powering up
the MiniCAT™ scanner hardware or workstation/server.
Installation/Relocation
The MiniCAT™ cannot be installed or relocated by the user. Proper and safe setup requires
specialized training and must be performed by a Xoran Technologies Customer Service Engineer.
Call Xoran 1-800-70-XORAN (800-709-6726) for assistance.