Secondary dose measurements per scan 0.5 meters below central slice at
radius of 1 meter
Azimuthal Location
Dose Per Scan (millirad)
A
1.07
B
0.93
C
0.98
D
0.84
E
0.91
F
0.94
Secondary dose measurements per scan 0.5 meters above central slice at
radius of 1 meter
Azimuthal Location
Dose Per Scan (millirad)
A
0.60
B
0.70
C
0.76
D
0.02
E
0.78
F
1.76
10.5 Dose Information
SourceRay Source
The CT conditions of operation used to provide dose and imaging performance information below
are as follows:
Sinus 20s (600): 120 kVp, 7 mA, 11.5 ms pulses, 600 pulses (48.3 mAs), 11.7 mm Al equivalent
fixed filtration, 0.542 ± 0.048 mm tomographic section thickness.
Sinus 10s (300): 120 kVp, 7 mA, 11.5 ms pulses, 300 pulses (24.15 mAs), 11.7 mm Al equivalent