MiniCAT™ User Manual v5.2
© 2014 Xoran Technologies LLC
10.4 Scatter Map
Secondary Radiation Measurements
Secondary scatter radiation was measured on the MiniCAT™ at 6 azimuthal locations A through F
and at radii 1.0, 1.5, and 2.0 meters from isocenter and in the same plane as the central slice (i.e.
level with the focal spot) which was approximately 120 centimeters above the floor (see diagram
below).
The MiniCAT™ was operated at 120 kVp and 48.3 mAs per Sinus 20s (600) protocol for all
secondary radiation measurements reported.