PR 5500 Instrument Manual
Device Installation
Sartorius EN-57
4.4.3 PR 5500/07 Analog Inputs and Outputs
The plug-in card for the available analog channels has 1 analog output (active) and 1 analog input.
The card is inserted in the Option 1/FB and/or Option 2 slot.
A max. of 2 PR 5500/07 cards can be installed!
PR 5500/07
AGND
V
in
0…10 V
I
out
0…20 mA
AGND
AGND
I
in
0…20 mA
10 E
D
D
A
A
DC
DC
50 K
25 K
200 E
AGND
+
DAC
ADC
Output
1 active current output:
0/4 to 20 mA (max. 24 mA),
10 V output voltage via an external
500 ¥ resistor
Output function
Gross/Net/Display follows or
application-dependent
Output range
0/4 to 20 mA, configurable
Output resolution
Internal 16 Bits = 65536 counts,
resolution of 20,000 @ 20 mA
Linearity error
output
@ 0 - 20 mA: 0.04 %
@ 4 - 20 mA: 0.02 %
Temperature error
output
<100 ppm/K
Zero point error
output
0.05 %
Max. error output
<0.1 %
Load
max. 0 to 500 ¥
Potential isolation
Yes
Cable
<150 m screened
Input
1 current or voltage input
Input range
0…10 V or 0…20 mA
Input resistance
100 k¥ for 10 V measuring input
200 ¥ for 20 mA measuring input
Input resolution
internal 14 Bits binary = 20,000
counts, @ e.g. 0 to 20 mA/0 to 10 V
Max. error input
0.2 %
Linearity error
input
<0.02 %
Temperature error
input
<75 ppm/K
Reserve input
±15%, i.e. –1.5 V to +11.5 V
Potenzialtrennung
Yes, joint inputs and outputs
Cable
<150 m screened
Intern. connection
Contact strip
Extern. connection
2x terminal, 6-pin
Dimens. (LxWxH)
50x45x18 mm
Weight
Approx. 40 g