PR 5500 Instrument Manual
Device Installation
Sartorius EN-37
4.3.5 RS-232 Interface (built-in)
The device is equipped with an integrated RS-232 interface. This interface is configurable, and can be used, for
example, for data transmission to a remote display or printer.
Connection
Terminal, 5-pin
Number of
channels
1
Type
RS-232, full duplex
Transmission rate
300 to 115k2 Bit/sec
Parity
None, odd, even
Data bits
7/8 bit
Input signal level
Logic 1 (high) - 3 to - 15 V
Logic 0 (low) + 3 to + 15 V
Output signal level
Logic 1 (high) - 5 V to - 15 V
Logic 0 (low) + 5 V to + 15 V
Number of signals
Input: Rx, CTS
Output: Tx, RTS
GND
CTS
RxD
TxD
RTS
Potential isolation
none
Cable gauge
max. 1.5 mm
2
Cable length
max. 15 m
Cable type
Twisted pair, screened
(e.g. LifYCY 3x2x0.20), 1 pair of
wires for ground (GND).
TxD
RxD
CTS
GND
RTS