PR 5500 Instrument Manual
Commissioning
Sartorius EN-163
5.11.7 Determining the Dead Load [Dead load at]
Note
If a liniarization was executed (see Chapter 5.11.12): After selection of the line "Dead load at"
the following tip is displayed:
"cannot be changed here
while linearisation is active"
Only deletion of linearization points deactivates linearization!
To use the empty scale/hopper as dead load (normal
case):
t
t
Clear the scale/hopper.
t
t
Press the [by load] soft key.
t
t
Confirm the entries.
t
y
The confirmation is displayed with "setting dead
load..."
If the mV/V value of the dead load was calculated, or
if it is known from the previous calibration, the value
can be overwritten by pressing [by mV/V].