4-9. Calibration only of ZERO
This function is for calibrating again only of the ZERO when the amount of the tare other
than the maximum capacity of the measuring part are changed.
1
2
3
Change to calibration mode only for ZERO
Step
1
Measurement mode
Calibration mode only for ZERO
By pushing key from the normal measurement mode,
displays.
4
The display becomes . By pushing key,
displays.
Push key at three times. The display changes as follows:
→
→
→
▲
1
2
3
Calibration only for ZERO
Step
2
Measurement mode
Push key to finish this mode.
By pushing key , blinks and the calibration of
ZERO starts.
▲
By pushing key from the display of ,
* Please set the condition that nothing is put on the measuring part.
The calibration of ZERO is finished, and displays
4
By pushing key twice, you can return to the measuring mode.
Display becomes , the set data is memorized
to the internal memory.
After lights on, push key.
STABLE
→
→
→
By pressing key three times, display changes in order as
Display becomes , you can enter into the calibration
mode only for ZERO.
displays.
F
ENTER
FUNC.
FUNC.
CAL-Z
CAL-Z
ENTER
ENTER
FF
CAL-Z
CAL-Z
ENTER
ENTER
FF
ENTER
ENTER
FF
ENTER
ENTER
FF
FUNC.
FUNC.
Error display of ZERO calibration
blinks for about 2 seconds when the load cell output value exceeds over ZERO adjustment
range at minus side (at -2.5 mV/V or more)
blinks for about 2 seconds when the load cell output exceeds over zero adjustment
range at plus side (at 2.5 mV/V or more), or the value of [Registered SPAN + ZERO
registering again] exceeds over 3.1 mV/V.
* Please con
fi
rm the rated capacity and the tare amount, etc., so that the ZERO should not exceed the adjustable
range.