Register 85
ResetPwf
Reset the PowerFail bit (Register 1; the bit was
set after power on).
Register 86
SetFixTare
Taring with weight in numerical address D 31
"Fixtare."
Register 87
GetFixTare
The current gross weight is copied into numeri-
cal address D31.
Register 89
ResetError
The
CmdError
error bit is reset.
Register 90
GetAlibi
Export selected data record from alibi memory.
Register 91
GetDate
Read current date and time.
Register 92
SetDate
Store (set) current date and time.
12.2.6.13
Registers 112…124: Transition-controlled action bits (Write)
For setting bits, see Chapter
As soon as the bit was set, it is reset internally and the operation is handled; this is
transition-controlled (for writing once).
The bit is set as
Write_Value_Select
with the specified number (see Chapter
).
Register 112
SetZero
Register 113
SetTare
Register 114
ResetTare
Register 115
SetTest
Register 116
ResetTest
Register 117
ResetPwf
Register 118
SetFixTare
Register 119
GetFixTare
Register 121
ResetError
Register 122
GetAlibi
Register 123
GetDate
Register 124
SetDate
Note:
To prevent excessively frequent writing in the EAROM memory, the write interval should
not be less than 15 seconds.
12.3 Filling protocol (64-Byte) for the "EasyFill" application
The "EasyFill" application uses a 64-byte interface.
The 8-byte scale protocol for the "Standard" application is not available.
12.3.1
Write window (input area)
Data are transferred from the master (PLC) to the slave (PR 5410) in this window.
12 Fieldbus interface
X3 Process Indicator PR 5410
EN-305