7.15.11
xBPI linearization
The measurement range for a straight can be optimized by setting the linearization
points. The following describes standard linearization.
Requirements:
-
The xBPI protocol has been selected (see Chapter
-
The "xBPI-scale" weighing point has been selected (see Chapter
-
The platform has been set up (see Chapter
-
In the menu [Weighing point A] - [xBPI-Scale] - [Setup] at [Configuration] -
[Weighing parameters] - [Confirming adjust.] was set to "manual".
-
The communication between the device and platform is active.
Procedure:
2.
Press the [Setup] softkey.
The parameters of the xBPI-scale are read into the device.
3.
Select [Calibration]- [Adjust with intern weight] with the cursor and confirm.
The deviation is displayed in the last line with increased resolution (10-fold).
The procedure is displayed by status messages in a row.
4. Press the [Accept] softkey.
The weight is displayed in high-resolution (10x).
The date are saved.
5.
Press the
softkey to exit the menu and to save the settings.
1.
Select
-[Weighing point]- [xBPI-Scale] and confirm.
X3 Process Indicator PR 5410
7 Commissioning
Minebea Intec
EN-192