-
In the menu [Weighing point A] - [xBPI-Scale] - [Setup] at [Configuration] -
[Weighing parameters] - [Confirming adjust.] was set to "manual".
-
The communication between the device and platform is active.
Procedure:
7.15.10
xBPI calibration with built-in weight
Requirements:
-
The xBPI protocol has been selected (see Chapter
-
The "xBPI-scale" weighing point has been selected (see Chapter
-
The platform has been set up (see Chapter
-
In the menu [Weighing point A] - [xBPI-Scale] - [Setup] at [Configuration] -
[Weighing parameters] - [Confirming adjust.] was set to "manual".
-
The communication between the device and platform is active.
Procedure:
1.
Select
-[Weighing point]- [xBPI-Scale] and confirm.
2.
Press the [Setup] softkey.
The parameters of the xBPI-scale are read into the device.
3.
Select [Calibration]- [Adjust with default weight] with the cursor and confirm.
The weight is specified automatically.
The calibration process is carried out without a weight. The calibration status is
displayed.
4. Place the displayed weight on the scale.
The deviation is displayed in the last line with increased resolution (10-fold).
5.
Press the [Accept] softkey.
The weight is displayed in high-resolution (10x).
The date are saved.
6. Remove the weight.
7.
Press the
softkey to exit the menu and to save the settings.
1.
Select
-[Weighing point]- [xBPI-Scale] and confirm.
7 Commissioning
X3 Process Indicator PR 5410
EN-191
Minebea Intec