— D
The readability (1 position) can only be changed
via the
key or the
key.
— CAL.WGT
Calibration weight, which is used in [CAL.ADJ]-
[CAL]- [WP2] and can be freely selected within
30% of the max. load and min. load
— MAX.CAP
Max. load of the scale
If this entered load has been reached, "H" is
displayed.
— END
Menüebene verlassen
— SER.NUM
Display serial number (if two scales are
selected, two serial numbers are displayed)
— W1.471
Using the
(No) key or the
key
will display the next six digits or display the first
six digits again.
— W2.471
Using the
(No) key or the
key
will display the next six digits or display the first
six digits again.
— END
— TYPE
Display type name of the scale
— SFT.VER
Display software version
— SR 2.64
— GEO.DAT
Display selected GEO area, which is valid for
both scales
— 12
Selected GEO area
— WP1
Info WP1
— D
Readability
— MAX
Max. load
— END
— WP2
Info WP2, if selected in the menu
— D
Accuracy
— MAX
Max. load
— END
— LFT.MOD
Display status of the legal metrology mode: ON
or OFF for
both
scales
— OFF
Default mode
Note:
-
The increment (readability) of the second scale must be bigger than the increment
of the first scale (d2 > d1).
Example: If the readability of the first scale = d1 = 1 g, then the readability of the
second scale must be = d2 = 2 g (d2 cannot be 1 g or 0.5 g).
-
The quotient from calibration weight and readability (CAL.WGT / d) must be ≥
1,000. This also affect the max. load (MAX.CAP), as: MAX.CAP ≥ CAL.WGT.
AD.CON2
5.3.4.10
[INFO] menu selection
INFO