5.2 Mounting without mounting kit
If the load cell is mounted without the mounting kit, the cell must be screwed onto a
corresponding holder in accordance with the load direction (arrow on the face of the load
cell).
The following fastening elements must be used:
Screw
Spring washer
Washer
Tightening torque
M8×35 ISO 4017-A2-70
A8 DIN 127-V2A
A8 ISO 7092-V2A 16.5 Nm
M8×35 ISO 4017-8.8
A8 DIN 127-8.8
A8 ISO 7092-8.8 24.0 Nm
If the load cell is in the hazardous area (no protection from entering, blows, falling solid
objects), the bellow must be protected by a cover. The bellow sheath must have free
space on all sides.
5.3 Mounting with PR 6007/00 mounting kit