The threaded bush (5) of the jack-up is cemented firmly into the lower plate (6).
3.3.2
Jack-up
WARNING
The vessel can tip over.
The threaded bolt (4) must only be turned upwards until the recess is visible.
Lift the vessel as follows:
Use the load cell as follows:
1.
Loosen the nut (8).
2.
Alternately turn the threaded bolt (4) upwards at the hex and turn the screw (2)
downwards to avoid jamming/wedging the upper plate (1).
3.
Repeat until the load cell is unloaded and can be removed together with the load
disc.
4. Insert the load disc into the load cell.
5.
Insert the load cell into the mounting kit.
6. Load the load cell slowly and adjust the built-in lift-off protection as described in
Chapter
3 Recommendations for installation
Maxi FLEX and Maxi FLEXLOCK mounting kits PR 6012/02, ../32, ../42
Minebea Intec
EN-9