320 … 1920 ms –> P53: 0
7.5.11
Parameter P43: Test operating mode
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
Byte 1
Byte 2
Byte 3
MODE
MODE 0 = absolute, 1 = relative, default = 0
7.5.12
Parameter P44: Standstill time
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
Byte 1
Byte 2
Byte 3
STIME
Time expressed in multiples of the measurement time at which a standstill is detected.
STIME: 1 … 9 times the measurement time, default = 1
7.5.13
Parameter P45: Standstill range
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
Byte 1
Byte 2
SRANGE MSB
Byte 3
SRANGE LSB
Limit to define the weight to be in standstill.
SRANGE: 0 … 10.00 d, default = 1 d
Value = d x 100
7.5.14
Parameter P46: Standstill timeout
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
Byte 1
Byte 2
Byte 3
TIME
Number of measurement times, during which the standstill condition should be reached.
TIME: 0 … 100 M (measurement times), default = 8
Transmitter Series PR 5211