Field
Size
Function
AboveMax
1 bit
The weight value has exceeded the max. (FSD), but is
still within max. + permissible overload (gross ≤ FS-
D+overload).
ADUError
1 bit
AD conversion error, see Chapter
.
7.2.2
Reading and writing data with function numbers
7.2.2.1
Reading data
Procedure:
Action of the master
Slave reaction
Write function number to
Read_Value_Select
.
Write requested data in
Read_Data
(bytes
0-3).
Copy
Read_Value_Select
to
Read_Value_Selected
.
Wait until
Read_Value_Selected
=
Read_Value_Select
.
Read requested data in
Read_Data
(bytes
0-3).
7.2.2.2
Writing data
Procedure:
1.
Write the function number as
Read_Value_Select
in byte 4 of the write window (e.g.,
9 = net weight).
2.
Wait until
Read_Value_Selected
in byte 4 of the read window is equal to
Read_Value_Select
of the write window.
The requested value is available in bytes 0-3.
1.
Wait until
Write_Active
= 0 in the read window (slave is ready to receive new data).
2.
Write value in bytes 0-3 of the write window.
3.
Write the function number as
Write_Value_Select
in byte 5 of the write window (e.g.,
"Basic” application: 190 = analog output 1).
4. Wait until
Write_Active
= 1 in the read window.
5.
Write 0 in byte 5 (
Write_Value_Select
).
Write_Active
is reset.
Transmitter Series PR 5211
7 ProfiBus interface
Minebea Intec
EN-64