![Minebea Intec Maxxis 5 Скачать руководство пользователя страница 134](http://html1.mh-extra.com/html/minebea-intec/maxxis-5/maxxis-5_original-installation-manual_1801172134.webp)
7.4.2
Electromagnetic Compatibility (EMC)
All data in compliance with NAMUR NE 21, EN 45501 and EN 61326.
High frequency electromagnetic ields
(80…3000 MHz)
EN 61000-4-3
10 V/m
Housing
Electrostatic discharge (ESD)
EN 61000-4-2
6/8 kV
Fast transients (burst)
EN 61000-4-4
1 kV
Peak voltages (surge) 1.2/50 µs
EN 61000-4-5
1/2 kV
Signal and control
lines
Conducted disturbances by high fre-
quency coupling (0.15…80 MHz)
EN 61000-4-6
10 V
Fast transient disturbances (Burst)
EN 61000-4-4
2 kV
Peak voltages (surge) 1.2/50 µs
EN 61000-4-5
1/2 V
Conducted disturbances by high fre-
quency coupling (0.15…80 MHz)
EN 61000-4-6
10 V
Voltage dips
EN 61000-4-11
0/40/70%
20/200/500ms
Mains inputs
Mains failure link
EN 61000-4-11
20 ms
7.4.3
RF interference suppression
Interference emission
EN 55011, Group 1, Limit class A, for industrial
sectors
7.5 Mechanics
7.5.1
Design
Control cabinet housing made of stainless steel
Blackbox housing made of stainless steel
Table-top housing made of stainless steel
Wall housing made of stainless steel
7.5.2
Dimensions
See Chapter
.
7.5.3
Weights
Type
Net weight
Shipping weight
Control cabinet housing
3.0 kg
Approx. 4.0 kg
Blackbox housing
3.0 kg
Approx. 4.0 kg
Table-top housing
5.7 kg
Approx. 6.7 kg
Wall housing
5.7 kg
Approx. 6.7 kg
Process Controller Maxxis 5 PR 5900
7 Technical data
EN-132
Minebea Intec
Содержание Maxxis 5