11.3.4
Arbitrary control of middle low
You can set up to 50 points for the open rate during middle low and the time that
reaches that open rate.
The open rate is a relative value to the full low position at the gate. The total time of each
point becomes the completing time of middle low.
The setting range of open rate of the gate is from [0.000 %] to [100.000 %] and the
setting range of the completing time of middle low is from [50 ms] to [10 000 ms].
Default setting:
When registering 3 points
Point 1: Open rate of the gate [80.000 %], Time [500 ms]
Point 2: Open rate of the gate [50.000 %], Time [500 ms]
Point 3: Open rate of the gate [20.000 %], Time [500 ms]
O
pen
rat
e
[%]
Time [ms]
Middle Flow
1
2
3
Full Flow
Note:
-
The open rate is a relative value to the full low position at the gate.
-
In this control method, the gate operates continuously by the constant speed until
reaching the open rate set before at the set time.
11.3.5
Quadratic control of middle low
The quadratic control can control the vibrational component by a rapid change in the low
rate by smoothly changing the low rate.
Inlection point C is located on the straight line that connects open rate A at the time of
completion of full low and open rate B at the time of starting of ine low. The control
characteristic can be changed by the set value of inlection point C. Time from A to B
becomes the completion time of middle low.
1. Select [MEDIUM FLOW] in middle low screen, and select [ARBITRARY], and then tap
[OK].
2. Select [ARBITRARY] in the middle low screen.
3. Select [NUMBER OF POINTS] in arbitrary control screen, and input the numbers of
points with ten keys, and then tap [OK].
4. Select [TIME] for each point, input the time to reach open rate with ten keys, and then
tap [OK].
5. Select [OPEN] for each point, and input the open rate of the gate. And then, tap [OK].
Weight Controller CSD-918
11 Weighing mode
EN-102
Minebea Intec