8
SBI Interface
You can enter certain commands to control weighing platform functions through the SBI
interface. Each command starts with an escape sequence. The command is always closed
with an end of command (EOC). The end of command can be any combination of CR and
LF. Every data after EOC and before ESC will be ignored by the scale.
Read displayed value:
ESC
P
EOC
Answer (16 Byte):
V
W
W
W
W
W
W
W
W
W
E
E
E
CR
LF
V
Sign
possible characters: "+", "–", " "
W
Weight Value
possible characters: "0"..."9", ". ", " "
E
Unit
possible characters: "a"..."z", "A"..."Z", " "
CR
Carriage Return
ASCII 0x0D
LF
Line Feed
ASCII 0x0A
Zero the weighing platform:
ESC
Z
EOC
Answer: see special Response Commands
Tare the weighing platform:
ESC
T
EOC
Answer: see special Response Commands
Special Response commands:
There are some special response commands. These Commands are used as Standard
responses. For example, error or confirmation. Special Response Commands always have
the same size of 5 Bytes!
OK
1
2
3
4
5
O
K
!
CR
LF
The Scale confirmed the command execution without any errors.
Compact Scale Puro® - Basic
8 SBI Interface
EN-36
Minebea Intec
Содержание Compact Scale Puro Advanced EF-P1
Страница 46: ...all dimensions in mm inch Compact Scale Puro Basic 12 Technical data EN 44 Minebea Intec...
Страница 48: ...LargeFlat LF Compact Scale Puro Basic 12 Technical data EN 46 Minebea Intec...
Страница 50: ...all dimensions in mm inch SmallTall ST Compact Scale Puro Basic 12 Technical data EN 48 Minebea Intec...
Страница 51: ...all dimensions in mm inch 12 Technical data Compact Scale Puro Basic Minebea Intec EN 49...
Страница 53: ...SmallFlatt SF all dimensions in mm inch 12 Technical data Compact Scale Puro Basic Minebea Intec EN 51...
Страница 55: ...all dimensions in mm inch 12 Technical data Compact Scale Puro Basic Minebea Intec EN 53...