D - 5
D.2.3
Temp Technical Alarm Messages
SpO2 No Cable
Connected
Low
C
No SpO
2
cable is connected. Connect
an SpO
2
cable.
Alarm message
Default
priority
Indication on
alarm reset
Cause and solution
Alarm message
Default
priority
Indication on
alarm reset
Cause and solution
Cannot Take
Neonatal Rectal
Temp (for
SmarTemp)
High
B
Rectal measurement is not
applicable for neonatal patient.
Select another site.
Temp No Probe
(for SmarTemp)
Med
B
The Temp probe is disconnected or
damaged. Reconnect the probe. If
the problem persists, change for
another probe.
Temp Probe
Misplaced (for
SmarTemp™)
Med
A
The measured value is invalid.
Replace the probe and check the
value.
Temp Measuring
Timeout (for
SmarTemp™)
Med
B
The measurement has exceeded
regular measurement time. Try
another measurement.
Ambient Temp
Overrange (for
SmarTemp™)
Low
A
The ambient temperature is too low.
Move the patient to a warmer place
and take temperature measurement
again if necessary.
Temp Overrange
(for
SmarTemp™)
Low
C
The temperature measurement
exceeds the measurement range.
Check the patient’s condition.
Thermal Resistor
Err (for
SmarTemp™)
Med
C
The Temp module may fail. Take
temperature measurement again. If
the alarm persists, contact your
service personnel.
TempIF
Thermometer
Error (for
Genius™ 3)
High
C
The thermometer may fail. Take
temperature measurement again. If
the alarm persists, replace the
thermometer.
Содержание VS 8A
Страница 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 8: ...2 6 This page intentionally left blank ...
Страница 20: ...12 D 2 9 Other System Technical Alarm Messages D 10 E Abbreviations E 1 F Declaration of Conformity F 1 ...
Страница 26: ...1 6 This page intentionally left blank ...
Страница 58: ...3 24 This page intentionally left blank ...
Страница 66: ...5 4 This page intentionally left blank ...
Страница 70: ...6 4 This page intentionally left blank ...
Страница 98: ...8 16 This page intentionally left blank ...
Страница 142: ...13 4 This page intentionally left blank ...
Страница 162: ...15 16 This page intentionally left blank ...
Страница 166: ...16 4 This page intentionally left blank ...
Страница 202: ...19 8 This page intentionally left blank ...
Страница 224: ...22 10 This page intentionally left blank ...
Страница 242: ...A 18 A 7 Operating Environment Host CPU AM3358 Primary programming language C Operating system Linux 3 2 0 ...
Страница 250: ...B 8 This page intentionally left blank ...
Страница 260: ...C 10 This page intentionally left blank ...
Страница 271: ...D 11 This page intentionally left blank ...
Страница 272: ...D 12 ...
Страница 275: ...F 1 F Declaration of Conformity ...
Страница 276: ...F 2 This page intentionally left blank ...
Страница 277: ......
Страница 278: ...P N 046 020097 00 4 0 ...