11 - 10
11.8.5
Setting the NIBP Display Format
To set the NIBP display format, follow this procedure:
1.
In
Continuous Monitoring
mode, select the NIBP numeric area to enter the
NIBP
menu. In
Spot Check
mode, select the
Main Menu
quick key
→
from the
Parameters
column select
Setup
→
select
NIBP
.
2.
Select the
Setup
tab.
3.
Set
Display Format
.
11.8.6
Setting the NIBP Alarm Limits Display Switch
In
Continuous Monitoring
mode, you can set whether to display the alarm limits of
diastolic NIBP and mean NIBP. Follow this procedure:
1.
Select the NIBP numeric area to enter the
NIBP
menu.
2.
Select the
Setup
tab.
3.
Switch on or off
Display Alarm Limits.
11.8.7
Selecting NIBP Measurement Algorithm
You can change the NIBP measurement algorithm adopted on your monitor. Follow this
procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required password
→
select
→
Module
.
2.
Under the
NIBP
tab, set
NIBP Measurement Algorithm
to:
◆
Inflation
: the monitor measures blood pressure during cuff inflation. If
succeeded, this measurement is completed. If the measurement fails at the
inflation stage, the monitor will automatically continue the measurement at
deflation stage.
◆
Deflation
: the monitor measures blood pressure during cuff deflation.
Affected by measurement factors or patient’s physiological conditions, inflation
algorithm might fail, be incorrect, or unavailable. Then the monitor automatically
continues to perform the NIBP measurement at deflation stage. The measurement
factors or patient’s physiological conditions include:
■
Excessive patient movement
■
Using cuffs with no
symbol
■
Using cuffs of inappropriate size
■
Cuffs being too tight or loose
■
Using well-worn cuffs
■
Patient blood pressure being too low
■
Patient pulse rate being too low
Содержание VS 8A
Страница 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 8: ...2 6 This page intentionally left blank ...
Страница 20: ...12 D 2 9 Other System Technical Alarm Messages D 10 E Abbreviations E 1 F Declaration of Conformity F 1 ...
Страница 26: ...1 6 This page intentionally left blank ...
Страница 58: ...3 24 This page intentionally left blank ...
Страница 66: ...5 4 This page intentionally left blank ...
Страница 70: ...6 4 This page intentionally left blank ...
Страница 98: ...8 16 This page intentionally left blank ...
Страница 142: ...13 4 This page intentionally left blank ...
Страница 162: ...15 16 This page intentionally left blank ...
Страница 166: ...16 4 This page intentionally left blank ...
Страница 202: ...19 8 This page intentionally left blank ...
Страница 224: ...22 10 This page intentionally left blank ...
Страница 242: ...A 18 A 7 Operating Environment Host CPU AM3358 Primary programming language C Operating system Linux 3 2 0 ...
Страница 250: ...B 8 This page intentionally left blank ...
Страница 260: ...C 10 This page intentionally left blank ...
Страница 271: ...D 11 This page intentionally left blank ...
Страница 272: ...D 12 ...
Страница 275: ...F 1 F Declaration of Conformity ...
Страница 276: ...F 2 This page intentionally left blank ...
Страница 277: ......
Страница 278: ...P N 046 020097 00 4 0 ...