System Maintenance 17-7
17.1.5 Checking Appearance
Check if there are any cracks in the covers:
Ultrasound system covers
Probe appearance.
External appearance of the ECG lead.
17.1.6 System Hard Drive Backup
To prevent deterioration or loss of data stored in the system hard drive (including patient info data,
preset data, etc.), create a backup copy of the hard drive at regular intervals.
17.2 Troubleshooting
If any persistent system malfunction is experienced, e.g., an onscreen error message, blank
imaging screen, absent menus, see the table below. If the failure cannot be resolved, contact the
Mindray Customer Service Department or a sales representative.
Troubleshooting Table
No.
Failure
Cause
Measure
1
The system cannot
be powered on.
Abnormal power system or
incorrect connection of the
power cord.
Check that the system is plugged in.
Check that the plug has not become
loose or dislodged from the back of
the system.
2
The display has no
output.
The interval between turning
off and restarting the system
is too short – wait at least 20
seconds.
Turn off the system and wait at least
1 minute, then restart the system.
The display brightness or
contrast may be improperly
set.
Adjust the display brightness and
contrast.
3
The touch screen
displays the
characters and
menus but no
images.
The acoustic power, gain or
TGC controls are improperly
set.
Adjust the [A.Power] control, gain or
TGC control.
Check that a probe is
connected and/or fully
connected.
Check probe connection.
The system is in frozen
status.
Unfreeze the image.
4
The image quality
is degraded
The exam mode is incorrect. Select an appropriate exam mode.
The image parameter
settings are incorrect.
Adjust the image parameters.
Содержание TE5
Страница 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Страница 2: ......
Страница 6: ......
Страница 12: ......
Страница 24: ......
Страница 36: ......
Страница 54: ......
Страница 56: ...4 2 Exam Preparation 4 1 1 New Patient Information The Patient Info screen is shown as follows 2 1 3 ...
Страница 110: ......
Страница 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Страница 120: ......
Страница 124: ......
Страница 156: ......
Страница 174: ......
Страница 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Страница 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Страница 203: ...Probes and Biopsy 13 3 7L4s P10 4s L20 5s P7 3s L14 5sp SC6 1s SP5 1s 6CV1s L9 3s C5 1s L11 3VNs C4 1s ...
Страница 222: ...13 22 Probes and Biopsy No Name Description 8 Grooves of the needle guided bracket Matched with the tabs of the probe ...
Страница 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Страница 250: ......
Страница 272: ......
Страница 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Страница 282: ......
Страница 318: ......
Страница 322: ......
Страница 323: ...P N 046 006959 07 1 0 ...