6 - 23
Test and Calibration
Recorder Check
6.11
Recorder Check
1.
Print SpO
2
Pleth waveforms. The recorder should print correctly and printout should be
clear.
2.
Open the recorder door and verify the monitor gives the proper message. Close
recorder door before proceeding to the next step.
3.
Set the recorder to print trend data. Check that the recorder prints accordingly.
Contact Technical Support if the recorder test fails.