Passport V Operating Instructions
3 - 11
Patient Management
Monitor/Display Troubleshooting
3.6
Monitor/Display Troubleshooting
MESSAGE/PROBLEM
REASON
SOLUTION
No trace for a desired
parameter
Improper attachment of
transducer or cable to monitor.
Check transducer / cable
connection.
Faulty transducer or cable.
Try a new transducer or cable.
Display Appears to be Off
Mains power switch may not be
on.
Press mains power switch on
side panel.
Unit may not be plugged into an
AC outlet.
Check power cord (Is it plugged
in
?
)
If used as a portable, battery
pack may be drained.
If battery pack is drained, plug
into an AC outlet to recharge the
battery. A period of 5.5 hours
(monitor OFF) or 8.5 hours
(monitor ON) is required for a
full charge of lithium-ion
batteries.
Monitor or display is damaged.
Contact Customer Support.
Disabled Alarm Tone
Silence
key pressed.
Check for alarm silence symbol
and message.
Beep volume low.
Increase beep volume.
Cooling Fan Failure
The unit running on AC power
and the cooling fan is not
operational.
Contact Customer Support.
Patient Information did not
appear on display
No data entered.
Enter proper patient data.
Done
was not selected from
keypad after entering data.
Go to the proper keypad enter
data, select
Done
when
finished.
Incorrect Date or Time
Data not entered or entered
incorrectly.
Follow instructions from
to Set the Clock / Date and
Time”
.
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...