SpO
2
Pulse Oximetry
SpO2 Monitoring
6 - 6
Passport V Operating Instructions
•
•
CAUTION:
CAUTION:
CAUTION:
CAUTION:
6.4.1
Masimo SET
®
SpO
2
Passport V
monitors equipped with Masimo SET SpO
2
allow the user to adjust
Sensitivity
and
SpO
2
Average Time
. The user should choose the sensitivity mode
depending upon signal quality and patient motion. In most cases, the normal setting is
appropriate. If patient motion is limited, high sensitivity can be used.
It is also possible to change the averaging time of the
Saturation
and
Pulse Rate
measurements. The
SpO
2
Average Time
can be changed to
2-4
,
4-6, 8
,
10
,
12
,
14
,
or 16 seconds
.
CAUTION:
Masimo SET Sensors
Masimo SET provides a family of sensors suitable for a wide variety of clinical settings and
patients. Specific sensors have been developed for neonates, infants, children, and adults.
All sensors are indicated for continuous non-invasive monitoring of arterial oxygen saturation
(SpO
2
) and pulse rate, all sensors are non-sterile and usable during patient movement.
The Adult Reusable Finger Sensors can also be used for spot check applications if needed.
All sensors are intended for single-patient use only unless indicated as “reusable.”
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...