Monitoring PR
10.1
Introduction
The pulse numeric counts the arterial pulsations that result from the mechanical activity of the heart. You can display a
pulse from any measured SpO
2
or any arterial pressure (see the IBP section). The displayed pulse numeric is color-coded
to match its source.
1. PR Source
2. PR: detected beats per minute.
10.2
Setting the PR Source
The current pulse source is displayed in the PR parameter area. The pulse rate chosen as pulse source:
is monitored as system pulse and generates alarms when you select PR as the active alarm source;
is stored in the monitor’s database and reviewed in the graphic/tabular trends; in trend graphs, as the PR curve is in
the same color with the PR source , it is unlikely to distinguish the PR source ;
is sent via the network to the central monitoring system, if available.
To set which pulse rate as PR source:
1. Enter the [
SpO
2
Setup]
menu.
2. Select [
PR Source
] and then select a label or [
Auto
] from the popup menu.
The popup menu displays the currently available PR sources from top to bottom by priority. When you select [
Auto
], the
system will automatically select the first option as the PR source from the popup menu. When the current PR source is
unavailable, the system will automatically switch [
PR Source
] to [
Auto
]. When you select [
IBP
], the system will
automatically select the first pressure label as the PR source from the popup menu.
2
1
Содержание iMEC8 Vet
Страница 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Страница 2: ......
Страница 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Страница 54: ...6 6 FOR YOUR NOTES ...
Страница 94: ...11 8 FOR YOUR NOTES ...
Страница 170: ...A 14 FOR YOUR NOTES ...
Страница 176: ...B 6 FOR YOUR NOTES ...
Страница 201: ......
Страница 202: ...PN 046 008143 00 3 0 ...