7-10
7.8
Silencing the Alarm Sound
You can silence all alarm sounds by pressing the
hardkey on the monitor’s front. In that case, the alarm lamp flashing
and alarm tones are cleared and
appears in the sound symbol area. After the physiological alarm is silenced, “
√
”
appears before the alarm message and the numeric and alarm limit still flash. For the performance after the technical
alarm is silenced, please refer to the
Silencing Technical Alarms
section.
The alarm silenced status will be automatically cancelled if you switch the monitor to other alarm statues or when a new
physiological or technical alarm occurs.
7.9
Latching Alarms
The alarm latching setting for your monitor defines how the alarm indicators behave when you do not acknowledge
them. When alarms are set to non-latching, their alarm indications end when the alarm condition ends. If you switch
alarm latching on, all visual and audible alarm indications last until you acknowledge the alarms, except that the
measurement numeric and violated alarm limit stop flashing as soon as the initial alarm condition goes away.
To set alarms to latching or non-latching:
1. Select [
Main Menu
]
→
[
Maintenance >>
]
→
[
User Maintenance >>
]
→
enter the required password.
2. Select [
Alarm Setup
>>].
3. Select [
Latching Alarms
] and toggle between [
High only
], [
Hi&Med
], [
All
] and [
Off
]. If you select [
High only
], only
high priority alarms are latched; if you select [
Hi&Med
], both high priority alarms and medium priority alarms are
latched; if you select [
All
], all alarms are latched; if you select [
Off
], the alarm latching is turned off.
NOTE
Changing of alarm priority may affect the latching status of corresponding alarms. Please determine if you
need to reset the latching status for the specific alarm when you have changed its alarm priority.
7.10
Silencing Technical Alarms
For some technical alarms, their alarm lamp flashing and alarm tones are cleared and the alarm messages change to
prompt messages after the
hardkey is pressed. After the monitor restores the normal status, the monitor can give
alarm indications correctly when these alarms are triggered again.
For some technical alarms, all their alarm indications are cleared after the
hardkey is pressed. After the monitor
restores the normal status, the monitor can give alarm indications correctly when these alarms are triggered again.
For some other technical alarms, their alarm lamp flashing and alarm tones are cleared and “
√
” appears before the alarm
message after the
hardkey is pressed. After the monitor restores the normal status, the monitor can give alarm
indications correctly when these alarms are triggered again.
Содержание iMEC8 Vet
Страница 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Страница 2: ......
Страница 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Страница 54: ...6 6 FOR YOUR NOTES ...
Страница 94: ...11 8 FOR YOUR NOTES ...
Страница 170: ...A 14 FOR YOUR NOTES ...
Страница 176: ...B 6 FOR YOUR NOTES ...
Страница 201: ......
Страница 202: ...PN 046 008143 00 3 0 ...