15-3
15.2.2
Making a Microstream CO
2
Measurement
Connect the sampling line to the monitor and then connect the CO
2
components as shown below. After warm-up is
finished, you can perform CO
2
measurements
15.3
Changing CO
2
Settings
15.3.1
Accessing CO
2
Menus
By selecting the CO
2
parameter window, you can access the [
CO
2
Setup
] menu.
15.3.2
Entering the Standby Mode
The standby mode of the CO
2
module relates to the standby mode of the monitor as follows:
If the monitor enters the standby mode, the CO
2
module will also enter the standby mode.
If the monitor exits the standby mode, the CO
2
module will also exit the standby mode.
If the CO
2
module enters or exits the standby mode, it will not affect the monitor.
To enter or exit the standby mode manually, select [
Operating Mode
] in the [
CO
2
Setup
] menu and then toggle
between [
Standby
] and [
Measure
].
When you set the sidestream CO
2
module to the strandby mode, the CO
2
gas sample intake pump automatically sets the
sample flow rate to zero. When exiting the standby mode, the CO
2
module continues to work at the preset sample flow
rate.
For the sidestream CO
2
module, you can set the delay time. After the delay time the CO
2
module enters the standby
mode if no breath is detected.
Sampling line
Connector for sampling line
Содержание iMEC8 Vet
Страница 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Страница 2: ......
Страница 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Страница 54: ...6 6 FOR YOUR NOTES ...
Страница 94: ...11 8 FOR YOUR NOTES ...
Страница 170: ...A 14 FOR YOUR NOTES ...
Страница 176: ...B 6 FOR YOUR NOTES ...
Страница 201: ......
Страница 202: ...PN 046 008143 00 3 0 ...