H - 2
Operator’s Manual
H Ultrasound Gel Heater
–
Ambient temperature: 0°C~40°C
–
Relative humidity: 20%~85% (no condensation)
–
Atmospheric pressure: 700hPa~1060hPa
•
Storage and transportation conditions
–
Ambient temperature: -20°C~55°C
–
Relative humidity: 20%~95% (no condensation)
–
Atmospheric pressure: 700hPa~1060hPa
H.3
Function and Requirement
NOTE:
•
When the ambient temperature is higher than the required temperature of the heater, the heater
does not function.
•
The heater can only heat one bottle of gel at a time.
The gel heater can make the gel reach 40°C. There are four levels for the temperature: 34°C, 37°C,
40°C and off.
When the gel is placed inside the heater, the time it takes to heat from an ambient temperature of
18°C to the desired temperature should be no more than 0.5 hours.
The ultrasound gel heater can work continuously over 12 hours.
H.4
Install the Heater
Perform the following procedure:
1.
Push the gel heater into the gel heater bracket.
2.
Plug the gel heater into the slot lying beneath the control panel.
3.
Put the gel inside the heater, and press the power button of the gel heater.
Содержание Imagyn 7
Страница 2: ......
Страница 14: ...This page intentionally left blank...
Страница 20: ...This page intentionally left blank...
Страница 54: ...This page intentionally left blank...
Страница 72: ...This page intentionally left blank...
Страница 118: ...This page intentionally left blank...
Страница 126: ...This page intentionally left blank...
Страница 196: ...This page intentionally left blank...
Страница 240: ...This page intentionally left blank...
Страница 280: ...This page intentionally left blank...
Страница 298: ...This page intentionally left blank...
Страница 406: ...This page intentionally left blank...
Страница 416: ...This page intentionally left blank...
Страница 491: ......
Страница 492: ...P N 046 019593 01 3 0...