9 - 3
•
Lethal arrhythmia alarms, apnea, and SpO
2
Desat are exclusive high priority alarms. When these
alarms occur, the monitor only displays messages of exclusive alarms. Other high priority alarms will
not be displayed. When multiple exclusive alarms occur simultaneously, alarm messages are
displayed circularly.
9.3.4
Alarm Status Symbols
Apart from the alarm indicators as described in
, the monitor uses the following symbols
to indicate the alarm status:
9.4
Accessing On-screen Help for Technical Alarms (AlarmSight)
In the technical alarm list, alarm messages followed by
Detail
include help messages or pictures to help you
identify the problem. This function is called AlarmSight. To access AlarmSight, follow this procedure:
1.
Select the technical alarm information area to enter the
Alarms
window.
2.
Select the
Technical Alarms
tab.
3.
From the alarm list select the desired alarm.
9.5
Checking Physiological Alarm List
To check the physiological alarm list, follow this procedure:
1.
Select the physiological alarm information area to enter the
Alarms
window.
2.
Select the
Physiological Alarms
tab.
9.6
Changing Alarm Settings
Select the
Alarm Setup
quick key or from the
Alarm
column of the main menu select desired buttons to set
alarm properties.
9.6.1
Setting Parameter Alarm Properties
To set parameter alarm properties, follow this procedure:
1.
Access the
Limits
page in either of the following ways:
◆
Select the
Alarm Setup
quick key.
◆
Select the
Main Menu
quick key
→
from the
Alarm
column select
Limits.
2.
Select a parameter tab and set alarm properties as desired. Enter the password if required. For more infor-
mation, see
27.12 The Authorization Setup Settings
You can also change the alarm properties of individual parameter from corresponding parameter menu.
Alarm pause:
indicates that all the alarms are paused.
Alarm off:
indicates that individual measurement alarms are turned off or the system is in the
alarm off status.
Audio pause:
indicates that audible alarm tones are paused.
Audio off:
indicates that audible alarm tones are turned off.
Alarm reset:
indicates that the alarm system is reset.
Содержание ePM 10M
Страница 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 58: ...4 8 This page intentionally left blank ...
Страница 62: ...5 4 This page intentionally left blank ...
Страница 118: ...11 4 This page intentionally left blank ...
Страница 134: ...13 12 This page intentionally left blank ...
Страница 144: ...15 8 This page intentionally left blank ...
Страница 156: ...16 12 This page intentionally left blank ...
Страница 174: ...18 12 This page intentionally left blank ...
Страница 182: ...19 8 This page intentionally left blank ...
Страница 192: ...20 10 This page intentionally left blank ...
Страница 222: ...24 4 This page intentionally left blank ...
Страница 228: ...25 6 This page intentionally left blank ...
Страница 256: ...28 6 This page intentionally left blank ...
Страница 264: ...29 8 This page intentionally left blank ...
Страница 268: ...30 4 This page intentionally left blank ...
Страница 280: ...31 12 This page intentionally left blank ...
Страница 346: ...E 4 This page intentionally left blank ...
Страница 350: ...F 4 This page intentionally left blank ...
Страница 360: ...G 10 This page intentionally left blank ...
Страница 361: ...H 1 H Declaration of Conformity ...
Страница 362: ...H 2 This page intentionally left blank ...
Страница 363: ......
Страница 364: ...P N 046 012607 00 6 0 ...