3 - 16
3.12.4
Printing Frozen Waveforms
To print the frozen waveforms, select the
button at the upper left corner of the
Freeze
window.
3.13
Capturing the Screen
The monitor provides the function of screen capture. To capture the current screen display, follow this
procedure:
1.
Connect the USB drive to the monitor’s USB connector.
2.
Press and hold the
More
quick key. Wait till it turns from blue to grey.
The captured pictures are automatically saved in the USB drive.
3.14
Checking Software Licenses
To run the following functions in your monitor, software licenses are required:
■
Early Warning Score (EWS)
■
ECG 24h Summary
To check the licenses, select the
Main Menu
quick key
→
select
License
→
Local
.
To install the licenses, follow this procedure:
1.
Connect the USB drive with the licenses in to the monitor’s USB connector.
2.
Select the
Main Menu
quick key
→
select
License
→
select
External
.
3.
Select
Install
.
3.15
Turning Off the Monitor
Before turn off the monitor, perform the following check:
1.
Ensure that the monitoring of the patient has been completed.
2.
Disconnect the cables and sensors from the patient.
3.
Make sure to save or clear the patient monitoring data as required.
To turn off the monitor, press and hold the power switch for 3 seconds.
Turning off the monitor does not disconnect the monitor from the AC mains. To completely disconnect the
power supply, unplug the power cord.
CAUTION
•
Press and hold the power switch for 10 seconds to forcibly shut down the monitor if it could not be
shut down normally. This may cause loss of patient data.
NOTE
•
The monitor that was switched on prior to a power loss automatically switched on when the power is
restored.
•
In case of a temporary power failure, if the power is restored within 30 minutes, monitoring will
resume with all active settings unchanged; if the monitor is without power for more than 30
minutes, the monitor behaves the same as it is normally turned off.
Содержание ePM 10M
Страница 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 58: ...4 8 This page intentionally left blank ...
Страница 62: ...5 4 This page intentionally left blank ...
Страница 118: ...11 4 This page intentionally left blank ...
Страница 134: ...13 12 This page intentionally left blank ...
Страница 144: ...15 8 This page intentionally left blank ...
Страница 156: ...16 12 This page intentionally left blank ...
Страница 174: ...18 12 This page intentionally left blank ...
Страница 182: ...19 8 This page intentionally left blank ...
Страница 192: ...20 10 This page intentionally left blank ...
Страница 222: ...24 4 This page intentionally left blank ...
Страница 228: ...25 6 This page intentionally left blank ...
Страница 256: ...28 6 This page intentionally left blank ...
Страница 264: ...29 8 This page intentionally left blank ...
Страница 268: ...30 4 This page intentionally left blank ...
Страница 280: ...31 12 This page intentionally left blank ...
Страница 346: ...E 4 This page intentionally left blank ...
Страница 350: ...F 4 This page intentionally left blank ...
Страница 360: ...G 10 This page intentionally left blank ...
Страница 361: ...H 1 H Declaration of Conformity ...
Страница 362: ...H 2 This page intentionally left blank ...
Страница 363: ......
Страница 364: ...P N 046 012607 00 6 0 ...