19 - 6
19.9.4
Setting Auto Standby
The monitor enters the
standby mode automatically after the configured period of time if no breath is detected
since the last detected breath. To set the auto standby, follow this procedure:
1.
Select the AG numeric area or waveform area to enter the
Gas
menu.
2.
Select the desired gas tab.
3.
Select the
Setup
tab.
4.
Set
Auto Standby
.
19.9.5
Setting the Gas Waveform
To set the gas waveform, follow this procedure:
1.
Select the AG numeric area or waveform area to enter the
Gas
menu.
2.
Select the desired gas tab.
3.
Select the
Setup
tab.
4.
Set the speed and scale of gas waveforms. For CO
2
, you can also set
Waveform Type
.
19.9.6
Setting the RR Source
To set the RR (respiration rate) source, follow this procedure:
1.
Select the AG numeric area or waveform area to enter the
Gas
menu.
2.
Select the desired gas tab.
3.
Select the
Setup
tab.
4.
Set
RR Source
.
When the current RR source does not have valid measurement, the system will automatically switch
RR Source
to
Auto
.
19.9.7
Entering the Intubation Mode
When performing intubation during general anesthesia, you can enter the intubation mode in order to reduce
unnecessary alarms. To enter the intubation mode, follow this procedure:
1.
Select the AG numeric area or waveform area to enter the
Gas
menu.
2.
Select
Intubation Mode
from the bottom of the menu.
For the details of the intubation mode, see
19.9.8
Enabling or Disabling MAC Display
You can set whether MAC value is displayed in the AG numeric area. To do so, follow this procedure:
1.
Select the AG numeric area or waveform area to enter the
Gas
menu.
2.
Select the desired anesthetic agent tab.
3.
Switch on or off
MAC
.
19.10 Changing the Anesthetic Agent
When the anesthetic agent used on the patient is changed, the AG module detects the mixed anesthetic gas
during the transition of two anesthetic agents. The time required for completing the replacement of anesthetic
agent depends on anesthesia type (low flow or high flow) and the characteristics of anesthetic agents
(pharmacokinetics). During the transition of two anesthetic agents, the monitor gives no prompt messages and
the MAC value displayed may be inaccurate.
Содержание ePM 10M
Страница 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 58: ...4 8 This page intentionally left blank ...
Страница 62: ...5 4 This page intentionally left blank ...
Страница 118: ...11 4 This page intentionally left blank ...
Страница 134: ...13 12 This page intentionally left blank ...
Страница 144: ...15 8 This page intentionally left blank ...
Страница 156: ...16 12 This page intentionally left blank ...
Страница 174: ...18 12 This page intentionally left blank ...
Страница 182: ...19 8 This page intentionally left blank ...
Страница 192: ...20 10 This page intentionally left blank ...
Страница 222: ...24 4 This page intentionally left blank ...
Страница 228: ...25 6 This page intentionally left blank ...
Страница 256: ...28 6 This page intentionally left blank ...
Страница 264: ...29 8 This page intentionally left blank ...
Страница 268: ...30 4 This page intentionally left blank ...
Страница 280: ...31 12 This page intentionally left blank ...
Страница 346: ...E 4 This page intentionally left blank ...
Страница 350: ...F 4 This page intentionally left blank ...
Страница 360: ...G 10 This page intentionally left blank ...
Страница 361: ...H 1 H Declaration of Conformity ...
Страница 362: ...H 2 This page intentionally left blank ...
Страница 363: ......
Страница 364: ...P N 046 012607 00 6 0 ...