10 - 17
10.7.5.3
Arrhythmia Alarm Shielding Rules
The following table explains how auidble and visual alarm indicate during arrhythmia alarm shielding period.
10.8
ST Segment Monitoring
ST monitoring is intended for adult, pediatric and neonatal patients.
10.8.1
ST Safety Information
WARNING
•
ST values may be affected by such factors as some drugs or metabolic and conduction disturbances.
•
ST deviation is often calculated at a fixed offset from the J point. Changes in heart rate may affect ST.
•
The ST deviation measurement algorithm has been tested for accuracy. The significance of ST
segment changes needs to be determined by a physician.
•
This monitor provides ST deviation level change information. The clinical significance of the ST level
change information should be determined by a physician.
10.8.2
Enabling ST Monitoring
The ST monitoring function is disabled by default. Before you start ST monitoring, enable the ST function. To do
so, follow this procedure:
1.
Select the ECG numeric area or waveform area to enter the
ECG
menu.
2.
Select the
ST
tab
→
select the
Setup
tab.
3.
Switch on
ST Analysis
.
Reliable ST monitoring cannot be ensured under the following situations:
■
You are unable to get a lead that is not noisy.
■
Arrhythmias, such as atrial fib or flutter, cause irregular baseline.
■
The patient is continuously ventricularly paced.
■
The patient has left bundle branch block.
In these cases, you may consider switching off ST monitoring.
Previous alarm
Current alarm
Alarm indication
Alarm in high priority
chain
Alarm in high priority chain
Alarm light and alarm tone
Alarm in medium priority chain
During the shielding period, alarm light and alarm
tone are disabled. When the shielding period is
reached, alarm light and alarm tone are
reactivated.
Alarm in medium
priority chain
Alarm in high priority chain
Alarm light and alarm tone
Alarm in the same medium priority
chain, but with higher priority
Alarm light and alarm tone
The same alarm reoccurs
During the shielding period, alarm light and alarm
tone are disabled. When the shielding period is
reached, alarm light and alarm tone are
reactivated.
Alarm in the same medium priority
chain, but with lower priority
During the shielding period, alarm light and alarm
tone are disabled. When the shielding period is
reached, alarm light and alarm tone are
reactivated.
Alarm in other medium priority chain
Alarm light and alarm tone
Содержание ePM 10M
Страница 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 58: ...4 8 This page intentionally left blank ...
Страница 62: ...5 4 This page intentionally left blank ...
Страница 118: ...11 4 This page intentionally left blank ...
Страница 134: ...13 12 This page intentionally left blank ...
Страница 144: ...15 8 This page intentionally left blank ...
Страница 156: ...16 12 This page intentionally left blank ...
Страница 174: ...18 12 This page intentionally left blank ...
Страница 182: ...19 8 This page intentionally left blank ...
Страница 192: ...20 10 This page intentionally left blank ...
Страница 222: ...24 4 This page intentionally left blank ...
Страница 228: ...25 6 This page intentionally left blank ...
Страница 256: ...28 6 This page intentionally left blank ...
Страница 264: ...29 8 This page intentionally left blank ...
Страница 268: ...30 4 This page intentionally left blank ...
Страница 280: ...31 12 This page intentionally left blank ...
Страница 346: ...E 4 This page intentionally left blank ...
Страница 350: ...F 4 This page intentionally left blank ...
Страница 360: ...G 10 This page intentionally left blank ...
Страница 361: ...H 1 H Declaration of Conformity ...
Страница 362: ...H 2 This page intentionally left blank ...
Страница 363: ......
Страница 364: ...P N 046 012607 00 6 0 ...