4.2.10
4.2.11
TSI
Description
Optimizes the image for different tissue:
Operations
Press <Menu> to open B Img Menu, and move cursor to [TSI (value)]. Press
<Set> or <Back> to adjust TSI.
General: It is helpful to scan the internal organs like liver, kidney, pancreas,
thyroid, etc.
Muscle: It is helpful to scan the muscle tissue or fibrous tissue, such as
muscle, ligaments, tendon, etc.
Fatty: It is helpful to scan tissue mainly composed of fat cell, such as lipoma,
breast of old women.
Fluid: It is helpful to scan tissue mainly consist of liquid, such as gallbladder,
cyst, vessels, etc.
Effects
/
Impacts
Adjusting TSI will clear the cine memory.
Operation
Conflicts
The TSI cannot be changed when the image is frozen.
Scan Mode
Scanning
Angle
Description
Adjusts the angle of scanning. It is available for B mode only, and the range is
0~3. 0 indicates the smallest scanning angle and 3 indicates the greatest scanning
angle.
Operations
Press <Menu> to open B Img Menu, and move cursor to [Scan Mode]
Æ
[Scan
Angle (value)]. Then press <Set> to increase the angle and press <Back> to
decrease the angle.
Effects
The larger the scan angle, the wider your field of vision. The less the scan angle,
the narrower your field of vision.
Impacts
Increasing the scanning angle will widen your field of vision while lowers the frame
rate. Decreasing the scanning mode will narrow your field of vision while increases
the frame rate.
Operation
Conflicts
The scanning angle cannot be changed when the image is frozen.
Scanning
Mode
Description
Changes the scanning mode.
Operations
Press <Menu> to open B Img Menu, and move cursor to [Scan Mode]
Æ
[Hi
Density1/Hi Density2/Hi Frm Rate]. Then press <Set> or <Back> to select the
scanning mode.
Effects
Select “Hi Density1/Hi Density2” to present clearer image and select “Hi Frm Rate”
to acquire higher frame rate.
“Hi Frm Rate” sacrifices image quality to acquire higher frame rate.
Impacts
Changing scanning mode will clear the cine memory.
Operation
Conflicts
The scanning mode cannot be changed when the image is frozen.
4-6 Optimizing the Image
Содержание DP-8300
Страница 1: ...DP 8500 DP 8300 Digital Ultrasonic Diagnostic Imaging System Operator s Manual Basic Volume ...
Страница 2: ......
Страница 8: ......
Страница 20: ......
Страница 36: ......
Страница 54: ......
Страница 58: ......
Страница 82: ......
Страница 86: ......
Страница 92: ......
Страница 94: ......
Страница 96: ......
Страница 101: ...P N 046 000308 00 V1 1 ...