Keyboard is abnormal
.
Is the problem resolved?
Replace the main board.
The main board's fault
The main panel's fault
.
Yes
No
Figure 5-47 Solution of the keyboard fault
4.3No Echo Signal in Image Area
Al l B mo de e ch o i ma ge s
are missing.
Replace the transducer
and use the same connector.
Connect different
transducer connector.
Is the problem resolved?
The transducer's
fault
The transducer
board's fault
Yes
Yes
No
No
Can the sound “chirp” be
heard from the
transducer?
No
Replace the
power board.
Is the problem
resolved?
PHV's fault
Yes
Yes
No
Is the problem resolved?
The main board's fault
yes
5-2
DP-3300/DP-3200
Service Manual
(