1 - 37
Theory of Operation
The Breathing System
1.9
The Breathing System
1.9.1
CMV mode, inspiration
Tidal volume (T Vol.) compensates for variations in gas flow. This is to ensure that the set tidal
volume is delivered to the patient.
FIGURE 1-28
Breathing System Pneumatics, CMV Mode, Inspiration
1.9.2
CMV mode, expiration
As the patient exhales tidal volume into the expiratory limb, fresh gas enters the bellows.
Fresh gas mixes with exhaled gas after the Absorber removes CO
2
. Excess fresh gas passes
through the exhaust valve to the AGSS
FIGURE 1-29
Breathing System Pneumatics, CMV Mode, Expiration
Breathing
System
Ventilator
Patient
Pressure
Regulator
Proportional
Valve
Solenoid
Valve
Fresh
G
a
s
PEEP
Valve
Exhaust
Valve
APL Valve
AGSS
Electronic
Control
Reversing
Valve
O2 Sensor
Insp. Valve
Absorber
Canister
Exp. Valve
Bre
a
th
in
g
Ba
g
P
Exp. Flow
Sensor
Insp. Flow
Sensor
Bellows
Breathing
System
Ventilator
Patient
Pressure
Regulator
Proportional
Valve
Solenoid
Valve
Fresh
G
a
s
PEEP
Valve
Exhaust
Valve
APL Valve
AGSS
Electronic
Control
Reversing
Valve
O2 Sensor
Insp. Valve
Absorber
Canister
Exp. Valve
Bre
a
th
in
g
Ba
g
P
Exp. Flow
Sensor
Insp. Flow
Sensor
Bellows
Содержание Datascope AS 3000
Страница 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Страница 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Страница 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Страница 127: ...Calibration Procedures Calibration 5 14 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Страница 178: ...AS3000 Service Manual 0070 10 0683 7 3 Warranty Manufacturer s Responsibility This page intentionally left blank...
Страница 179: ...0070 10 0683 Rev K August 8 2014...