6 - 31
Periodic Maintenance
Functional Tests
6.8.7.9
Pressure Support (PS) Ventilation Mode Test
1.
Attach a breathing circuit and breathing bag.
NOTE:
For testing purposes always use a reusable breathing
circuit.
2.
Attach an adult test lung to the Y-fitting of the breathing circuit.
3.
Attach a Vent Tester between the EXP port and the expiratory hose.
4.
Set the O
2
flow to 1 L/min and set the N
2
O and AIR flow rates to minimum flow.
5.
Set the ventilator controls to:
6.
Select
PS
to begin ventilation.
7.
Begin triggering breaths by slightly squeezing the test lung and releasing. Maintain a
continuous breath rate.
8.
Verify that a pressure waveform and all ventilation parameters appear on the screen.
9.
Verify that the Peak Pressure reading on the display is ±4 the value of
P + PEEP.
10.
Stop triggering breaths.
11.
Verify that after 30 seconds the ventilator delivers a breath and displays the message
APNEA BACKUP
.
12.
Verify the system ventilates with a frequency of 2 bpm
VENTILATOR CONTROLS
VENTILATOR SETTINGS
Vent Dial Patient Type
Adult
Vent Mode Ventilation Mode
PS
Differential Pressure -
P
20
PEEP -
PEEP
Off
Flow Trigger -
Trigger
3
Inspiratory Slope -
T
slope
0.5
Minimum frequency -
freq
MIN
2
Содержание Datascope AS 3000
Страница 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Страница 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Страница 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Страница 127: ...Calibration Procedures Calibration 5 14 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Страница 178: ...AS3000 Service Manual 0070 10 0683 7 3 Warranty Manufacturer s Responsibility This page intentionally left blank...
Страница 179: ...0070 10 0683 Rev K August 8 2014...