6 - 21
Periodic Maintenance
Functional Tests
3.
Select
Continue
to prompt to the Normal Screen.
FIGURE 6-21
Compliance Test Results
6.8.4.5
Manual Leak Test
NOTE:
The Manual Leak Test detects smaller leaks than can be
detected in the Automatic Leak Test.
1.
Ensure that the gas pressure for O
2
, N
2
O, and AIR are at 50 ±10 psi.
2.
Power ON the
3.
Attach a reusable-silicone-rubber breathing circuit to the Breathing System.
NOTE:
For testing purposes always use a reusable breathing
circuit.
4.
Tightly connect the Y-fitting on the breathing circuit to the test port.
5.
Attach a breathing bag to the bag arm.
6.
Set the APL Valve to the fully closed position (
70
cmH
2
O).
7.
Rotate the O
2
Flow Control Valve until 50 cmH
2
O pressure is observed on the Airway
Pressure Gauge.
8.
Verify that the flow required to stabilize the pressure is less than 300 mL/min.