BeneVision N1 Patient Monitor Operator’s Manual
6 - 7
3.
Select
Certification Management
from the bottom left corner of the menu.
◆
From the
Local
tab, select certifications you want to delete from the monitor, and then select
Delete
.
◆
From the
USB Drive
tab, select certifications you want to import from the USB memory, and then
select
Import
.
6.8.5
Setting Multicast Parameters
Multicast helps device discovery between monitors and between monitors and CMS. Devices in the same
multicast group can be mutually discovered.
To set multicast parameters, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required
password
→
select .
2.
Select the
Network
Setup
tab
→
Device Discover
tab.
3.
Set
Multicast TTL
.
4.
Set
Multicast Address
.
6.8.6
Setting the Network Service Quality Level
To set the quality of service (QoS), follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required
password
→
select .
2.
Select the
Network
Setup
tab
→
QoS
tab.
3.
Select
QoS Level For Realtime Monitoring
to set the service quality of network connection for realtime
monitoring, for example parameter measurements and waveforms, alarms, and so on. It is suggested that
you set the service quality for realtime monitoring to a higher level.
4.
Select
QoS Level For Others
to set the service quality of network connection for non-realtime monitoring,
for example history data, printing, and as on.
6.8.7
Selecting Data Encryption Type
To set data encryption type, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required
password
→
select .
2.
Select the
Network
Setup
tab
→
Encryption
tab.
3.
Select
Encryption Connection Type
to set data encryption type when connecting devices:
◆
Only Private Encryption
: Mindray private encryption is used to encrypt the transmitted data. You
cannot connect devices supporting SSL (secure sockets layer) encryption.
◆
SSL Encryption Priority
: for devices supporting SSL encryption, SSL encryption is used when
connecting the devices. For devices not supporting SSL encryption, private encryption is used when
connecting the devices.
6.8.8
Disabling the Broadcasting Patient Demographics Function
When viewing other patients, device location and patient information of remote devices are displayed in the
remote device list by default. To protect patient privacy, you can disable the broadcasting patient demographics
function to protect the monitor from sending patient information via broadcast. To do so, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required
password
→
select
.
2.
Select
Network Setup
tab
→
Information Security
tab.
3.
Switch off
Broadcast Patient Demographics
.
Thus, patient information will not display in the remote device list.