Monitoring ECG/Resp
7-13
7.6 About ST Monitoring
ST segment analysis is only for use on adult patients. The default setting is off.
ST segment analysis calculates ST segment elevations and depressions for individual
leads and then displays them as numerics in the ST1 and ST2 areas.
A positive value indicates ST segment elevation; a negative value indicates ST segment
depression.
Measurement unit of the ST segment: mV.
Measurement range of the ST segment: -2.0 mV to +2.0 mV.
7.6.1 Switching ST On and Off
To switch ST monitoring on or off:
1. In
the
[
ECG Setup
] menu, select [
ST Analysis >>
].
2. Select
[
ST Analysis
] to toggle between [
On
] and [
Off
].
When ST monitoring is switched on, the patient monitor selects the [
Diagnostic
] filter mode
automatically. You can also select the [
Monitor
] or [
Surgery
] filter mode. But in that case,
the ST numerics may be severely distorted.
By selecting the ST parameter window, you can enter the [
ST Analysis
] menu.
Содержание BeneView T8
Страница 1: ...BeneView T8 Patient Monitor Operation Manual...
Страница 2: ......
Страница 8: ...2 FOR YOUR NOTES...
Страница 24: ...Safety 1 6 FOR YOUR NOTES...
Страница 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Страница 74: ...Alarms 6 10 FOR YOUR NOTES...
Страница 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Страница 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Страница 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Страница 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Страница 184: ...Review 19 10 FOR YOUR NOTES...
Страница 206: ...Recording 21 8 FOR YOUR NOTES...
Страница 210: ...Other Functions 22 4 FOR YOUR NOTES...
Страница 256: ...Product Specifications A 24 FOR YOUR NOTES...
Страница 290: ...P N 6800 20 50657 1 0...