Monitoring Carbon Dioxide
13-5
13.3.3 Using a Mainstream CO
2
Module
1. Connect the sensor to the module.
2. By default, the mainstream CO
2
module is in standby mode. The message [
CO2
Standby
] appears on the screen when the CO
2
module is plugged. To set CO
2
module to
measure mode, select the
hardkey on the module to switch to the [
Measure
]
mode, or select the CO
2
parameter window
→
[
Operating Mode
]
→
[
Measure
]. The
message [
CO2 Sensor Warmup
] is displayed.
3. After warm-up is finished, connect the transducer to the airway adapter.
4. Perform a zero calibration per the
Zeroing the Sensor
section.
5. After the zero calibration is finished, connect the airway as shown below.
5. Make sure there are no leakages in the airway and then start a measurement.
NOTE
z
Always position the sensor with the adapter in an upright position to avoid
collection of fluids on the windows of the adapter. Large concentrations of fluids at
this point will obstruct gas analysis.
Airway adapter
Sensor
Connect to patient
Connect to monitor
Содержание BeneView T8
Страница 1: ...BeneView T8 Patient Monitor Operation Manual...
Страница 2: ......
Страница 8: ...2 FOR YOUR NOTES...
Страница 24: ...Safety 1 6 FOR YOUR NOTES...
Страница 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Страница 74: ...Alarms 6 10 FOR YOUR NOTES...
Страница 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Страница 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Страница 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Страница 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Страница 184: ...Review 19 10 FOR YOUR NOTES...
Страница 206: ...Recording 21 8 FOR YOUR NOTES...
Страница 210: ...Other Functions 22 4 FOR YOUR NOTES...
Страница 256: ...Product Specifications A 24 FOR YOUR NOTES...
Страница 290: ...P N 6800 20 50657 1 0...