Monitoring IBP
11-6
2. Vent the transducer to atmospheric pressure by turning on the stopcock to the air.
3. Press
the
hardkey on the module, or, in the setup menu for the pressure (e.g. Art),
select [
Art Zero >>
]
→
[
Zero
]. During zero calibration, the [
Zero
] button appears
dimmed. It recovers after the zero calibration is completed.
4. After the zero calibration is completed, close the stopcock to the air and open the
stopcock to the patient.
NOTE
z
Your hospital policy may recommend that the ICP transducer is zeroed less
frequently than other transducers.
11.7 Making the Pressure Calibration
A calibration ensures accurate pressure readings. Perform a calibration when you use a new
transducer, and at regular intervals according to your hospital policy. You require:
Standard sphygmomanometer
Balloon pump
Tubing
T-shape connector
WARNING
z
Never perform the invasive pressure calibration while a patient is being monitored.
3-way stopcock
Pressure transducer
Pressure adapter cable
Monitor
Содержание BeneView T8
Страница 1: ...BeneView T8 Patient Monitor Operation Manual...
Страница 2: ......
Страница 8: ...2 FOR YOUR NOTES...
Страница 24: ...Safety 1 6 FOR YOUR NOTES...
Страница 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Страница 74: ...Alarms 6 10 FOR YOUR NOTES...
Страница 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Страница 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Страница 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Страница 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Страница 184: ...Review 19 10 FOR YOUR NOTES...
Страница 206: ...Recording 21 8 FOR YOUR NOTES...
Страница 210: ...Other Functions 22 4 FOR YOUR NOTES...
Страница 256: ...Product Specifications A 24 FOR YOUR NOTES...
Страница 290: ...P N 6800 20 50657 1 0...