Monitoring IBP
11-2
11.3 Setting Up the Pressure Measurement
1. Plug the pressure cable into the IBP connector.
2. Prepare the flush solution.
3. Flush the system to exhaust all air from the tubing. Ensure that the transducer and
stopcocks are free of air bubbles.
WARNING
z
If air bubbles appear in the tubing system, flush the system with the infusion
solution again. Air bubble may lead to wrong pressure reading.
4. Connect the pressure line to the patient catheter.
5. Position the transducer so that it is level with the heart, approximately at the level of the
midaxillary line.
6. Select the appropriate label.
7. Zero the transducer. After a successful zeroing, turn off the stopcock to the atmospheric
pressure and turn on the stopcock to the patient.
Heparinized
fluid bag
Flushing set
Valve
Pressure
transducer
Stopcock
Adapter cable for using
disposable transducers
Monitor
Pressure line
Содержание BeneView T8
Страница 1: ...BeneView T8 Patient Monitor Operation Manual...
Страница 2: ......
Страница 8: ...2 FOR YOUR NOTES...
Страница 24: ...Safety 1 6 FOR YOUR NOTES...
Страница 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Страница 74: ...Alarms 6 10 FOR YOUR NOTES...
Страница 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Страница 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Страница 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Страница 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Страница 184: ...Review 19 10 FOR YOUR NOTES...
Страница 206: ...Recording 21 8 FOR YOUR NOTES...
Страница 210: ...Other Functions 22 4 FOR YOUR NOTES...
Страница 256: ...Product Specifications A 24 FOR YOUR NOTES...
Страница 290: ...P N 6800 20 50657 1 0...