A - 12
Defibrillator/Monitor Operator’s Manual
A.6
Power Supply Specifications
A.6.1
External Power Supply Specifications
A.6.2
Battery Specifications
Sample flowrate tolerance
±15% or ±15 ml/min, whichever is greater.
Rise time
Measured with a Oridion sampling line: ≤250 ms @ 50 ml/min (standard
sampling line) or ≤280 ms @ 50 ml/min (extended sampling line)
Response time
Measured with a Oridion sampling line: ≤5 s @ 50 ml/min (standard
sampling line) or ≤6.5 s @ 50 ml/min (extended sampling line)
awRR measurement range
0 to 150 rpm
awRR accuracy
<60 rpm:
60 to 150 rpm:
±1 rpm
±2 rpm
awRR resolution
1 rpm
Effect of interference gases on CO
2
measurements
Gas
Concentration (%)
Quantitive effect*
N
2
O
≤60
±1 mmHg
Hal
≤4
Sev
≤5
Iso
≤5
Enf
≤5
Des
≤15
±2 mmHg
*: means an extra error should be added in case of gas interference when CO
2
measurements are performed between 0-
40mmHg.
Inaccuracy specifications are affected by the breath rate and I: E change. The EtCO
2
accuracy is within specification for
breath rate ≤ 60rpm and I/E ratio ≤1:1, or breath rate ≤30rpm and I/E ratio ≤ 2:1.
AC power
Line voltage
100 to 240 VAC (±10%)
Current
1.8 to 0.8A
Frequency
50/60Hz (±3Hz)
DC Power (with an external DC/AC adapter)
Input voltage
12VDC
Power consumption
150W
Battery type
Smart lithium ion battery, rechargeable and free of maintenance, one battery can be
installed, two types of batteries can be configured
Battery LI24I005A: 14.4V, 5600mAh
Battery LI24I001A: 14.8V, 3000mAh
Battery LI24I005A charge
time
Charged by the equipment
connected to the AC power
Less than 3 hours to 90% and less than 4 hours to 100%
with equipment power off;
Less than 5 hours to 90% and less than 6 hours to 100%
with equipment power on.
Charged by the charger station
Less than 5 hours to 90% and less than 6 hours to
100%.
Содержание BeneHeart D2
Страница 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Страница 2: ......
Страница 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 204: ......
Страница 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Страница 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 207: ......
Страница 208: ...P N 046 010599 00 9 0...