6-4
CAUTION
If the patient experiences discomfort or danger at a higher air bubble filter
level, monitor the patient's physical conditions and select the actual
needed level. Measures should be taken instantly if any abnormal
condition occurs.
6.5 Accumulate Air
1. Select [
Main Menu
]→[
General Options
]→[
Accumulate air
].
2. Select [
Accumulate air
]: 0.1-4ml is adjustable.
6.6 Drip Setting
1. Select [
Main Menu
]→[
General Options
]→[
Drip setting
].
2. Select [
gtt/min
]→[
On
], and set the “Drip”. On infusion parameters setting
interface of Rate Mode and Time Mode, “ml/h” and “gtt/min” can be switched. If
[
Off
] is selected, “ml/h” and “gtt/min” can not be switched.
6.7 Key Lock Function
When locked, an
icon in the upper-right corner of the screen emerges. The
following are two ways for automatic lock and manual lock:
Automatic Lock:
1. Select [
Main Menu
]→[
General Options
]→[
Auto-lock Time
].
2. Select [
Auto-lock Time
]: Off, 1-5min. After a specific time is set during the
running state, and if there is no operation or high level alarm within the set
Auto-lock time, the key board will be auto-locked. [
Off
] indicates switching off the
function.
Manual Lock: In the running interface, under the unlocking condition, press and
hold (>3 seconds)
to lock the key board.
Note: If unlocking is needed, press and hold (>3 seconds)
to unlock, it is
automatically unlocked during the high level alarm.
6.8 Reminder Function
1. Select [
Main Menu
]→[
General Options
]→[
Reminder Time
].
2. Select [
Reminder Time
]: Off, 1-5min. [
Off
] indicates switching off the function.