5.7 Ramp Up/Down Mode
In Ramp Up/Down Mode, by setting the [
Ramp up time
] and [
Ramp down time
], the
infusion pump will increase the rate (divided as 9 stages) automatically during the
[
Ramp up time
]until the rate is steady, sustaining for a period of time, and then
decrease the rate during the [
Ramp down time
].
Mode
Parameters
Parameter Range
Ramp
Up/Down
Mode
VTBI
0.1-9999.99ml
Total time
00:00:01-99:59:59 h:m:s
Ramp up time
00:00:01-99:59:59 h:m:s
Ramp down time
00:00:01-99:59:59 h:m:s
NOTE:
1. [
Steady rate
], [
Up rate
], and [
Down rate
] are calculated automatically, and the
range is the same as the rate of Rate Mode.
2. When both ramp up time and ramp down time are null, the pump will run at the
Steady rate until the infusion is complete. Steady rate=VTBI/Total time.