A-3
A.3 Hardware Specifications
A.3.1 Display
Display
Type
Monochrome LCD
Size (diagonal)
3 inches
Differentiation
240
x
128 pixels
A.3.2 Battery
Internal battery
No. of batteries
1 (standard) or 2 (optional)
Battery type
Lithium battery
Shutdown delay
At least 30 mins (new battery, after the first low battery alarm)
Battery voltage
7.2V
Battery capacity
2600 mAh (1 battery) or 5200 mAh (2 batteries)
Power supply time
By factory default, continuously operate at a rate of 25ml/h,
discharge for at least 4h (1 battery) or 8h (2 batteries) using a
fully charged new battery.
Note: Power supply time is relevant to rate. By factory default,
continuously operate at the maximum selectable rate, typically
discharge for 1.5h (1 battery) or 4h (2 batteries) using a fully
charged new battery.
Charging time
When the pump is off, the charging time is not longer than 6h
(1 battery) or 12h (2 batteries).
A.3.3 Host LED
Host LED
Alarm light
1 (red/yellow bi-color)
AC/DC indicator light
1 (green)
Battery indicator light
1 (green)